CTS-1008plus Ultrasonic Flaw Detector

CTS-1008plus is a portable ultrasonic flaw detector that successfully makes a perfect combination of the advanced performance of normal pulse echo (PE) testing and TOFD imaging testing technology.

  • Product Intro
  • Feature
  • Specification
  • Scope

CTS-1008plus is a portable ultrasonic flaw detector that successfully makes a perfect combination of the advanced performance of normal pulse echo (PE) testing and TOFD imaging testing technology.

 

Its normal PE testing function supports the universal pulse-reflecting ultrasonic flaw detection.It has excellent performance, various powerful and useful functions. Besides, it supports various testing standards and application.

 

TOFD imaging technology refers to works using Time of Flight Diffraction technology and our composite material crystal transducer which effectively discover the defect easily and correctly. Meanwhile, CTS-1008Plus can be connected with many kinds of scanning devices(SCQcscan-1SCQtofd-1SCQtofd-2)and support B, C, D imaging modes.

 

It is only 1.4kg which makes it the lightest TOFD tester in China and completely meets your requirements on mobile testing and working under special testing circumstances. 。

  • TOFD imaging technology (More details please refer to the technical specifications).
  • Support TOFD recheck after normal PE testing, accurately measure the size and profile of defects, evaluate the characteristics and direction of defects. Support PE rechecks after TOFD imaging.
  • Sampling frequency up to 640MHz, 0.1mm measurement resolution, minimum display range is 2mm.
  • Frequency 0.5 ~ 20MHz, sensitivity up to 62dB, low frequency provides better signal-to-noise rate.
  • Bright, excellent resolution color TFT LCD to optimize the test result.
  • High performance square wave pulser with tuning option allows the operator to tune the pulse width of the square wave pulser to optimize transducer performance. This maximizes signal-to-noise and penetration in difficult materials such as highly-attenuated materials or thin work-pieces. +ve or –ve square wave or –ve square wave plus –ve spike pulse options.
  • Small size, lightweight, use Li-battery with more than 6 hours’ battery life.
  • LAN interface for data transfer and remote control.
  • Two-dimensional incremental encoder interface realizes accurate position-detecting imaging (B-scan, C-scan, D-scan).
  • Synchronous analogue RF waveform output with impedance 50 Ohm make it a good signal source of data collection and transducer test.
  • High-speed USB interface allows data storage on USB drive and PC communication. To interface with a mouse to secure accurate analysis on image.
  • Internal charger, auto switch between “charge” and “work”. Power supplied by battery or AC mains. The operating time of onboard Li battery is more than 6 hours.
  • Multi-software functions to meet all inspection needs, unique keypad organization and rotary knob operation make the inspection much easier. It has a user-friendly design for menu, English/Chinese language menu options and English/Chinese input options.
  • Automatically test the transducer frequency, tune the pulse width of square wave pulse to optimize transducer performance.
  • Thickness alarm function, continuous storage function of waveform and various measurement modes of waveforms.
  • Automatically calibrate instrument specifications and search the peak value.
  • Set-in API 5UE and AWS D1.1.
Pulse Type +ve square wave pulse, transmitter pulse voltage and width adjustable
Pulse Rise Time <10ns
Pulse Width Adjustable from 30 to 1000ns with 5ns increments
Transmitter Pulse Voltage 25~250V, continuously adjustable with 25V increments
Damping 400、80Ω
Pulse Repetition Frequency 25~1000Hz tunable
Rectification Mode Digital
Sampling Frequency 160MHz (Hardware)
Bandwith 0.5~20MHz (Broadband), 1~5MHz (Narrowband)
Dynamic Range >32dB
Linearity of Vertical Display <3%
Linearity of Time Base <0.2%
Resolutions >36dB
Surplus Sensitivity >62dB
Waveform Display Mode RF Echo
Material Velocity 1000~15000m/s
Range 0.0~10000mm longitudinal wave at steel velocity. Range continuously variable with minimum 0.1mm increments
Pulse Shift -7.5~3000μs
Triggering Mode Encoder Triggering (Max Resolution 0.125mm)
Equivalent Input Noise 20×10-9 V/√HZ
Accuracy of Calibrated Attenuator <±1dB/20dB
Imaging Mode B scan, D scan, TOFD imaging
Data Record Original Data Record
Scanning Length 5m each time, automatic scroll
Minimum Test Thickness <6mm using transducer we made to test
Scanner Various manual scanners optional
Transducer Composite material crystal
Data Analysis Recall A-scan waveform of scanned record
Defect size and profile recording and defect marking
Lateral wave straightening, backwall echo straightening and removing, contrast control, SAFT
Data Report Direct print of calibration table, A-scan waveform, B-scan image, D-scan image, TOFD image
I/O Port LAN, USB, two-dimensional incremental encoder
Control Mode Panel, rotary knob, mouse, keyboard
Outsourcable USB, Keyboard, Mouse, Printer with USB or LAN

.TOFD Transducer Specifications & Technical Data
  Specifications:

Frequency Crystal Diameter Connector Wedge
10MHZ 3mm L5(Microdot Equivalent)
Size of the transducer Casing: Approximately 22mm in height (including the connector)
Customization, regularly degree 45\60\70.
10MHz 6mm
5MHz 12mm
5MHz 6mm
2.5MHz 12mm
2.5MHz 6mm
1MHz 20mm

  Technical Data:(Take 5MHZ 6mm for example)

Transducer
Specification:TOFD 5MHz 6mm
Serial No.:II001
Testing Equipment
Pulse T/R Equipment::5077PR 04151704
Digital Oscilloscope:TDS3012B C011195
Cable:50Ω 2m
Test Condition
Pulse Voltage:100v
Transducer Frequency Options:5-6MHz
Gain:-24dB
Test Block:10mm PS
Test Result
-20dB Pulse Width:342.0ns
Central Frequency:5.2MHz
-6dB Relative Bandwidth:94.0%
Relative Sensitivity:-29.8dB

.TOFD Image Processing Technology (Minishing)

 I. (Synthetic Aperture Focusing Technique)SAFT

Pre-processed SAFT

Pro-processed SAFT

 II. Flatten-out

Pre Flatten-out

Pro Flatten-out

 III. Differentiation

Pre-differentiation

Pro-differentiation

.Normal Ultrasonic (PE) Testing Specifications

Pulse Type +ve square wave pulse, transmitter pulse voltage and width adjustable
Pulse Rise Time <10ns
Pulse Width Adjustable from 30 to 1000ns with 5ns increments
Transmitter Pulse Voltage 25~250V, continuously adjustable with 25V increments
Damping 400、80Ω
Test Modes Single/dual
Gain 0.0~110.0dB.Steps as 0.1,1.0,2.0,6.0dB
Pulse Repetition Frequency 25~1000Hz tunable
Rectification Mode Positive halfwave, negative halfwave, full wave, RF
Sampling Frequency 160MHz (Hardware)
Bandwith 0.5~20MHz(Broadband),1~5MHz(Narrowband)
Dynamic Range >32dB
Linearity of Vertical Display <3%
Linearity of Time Base <0.2%
Resolutions >36dB
Surplus Sensitivity >62dB
Material Velocity 500~15000m/s
Range 0.0~10000mm longitudinal wave at steel velocity. Range continuously variable with minimum 0.1mm increments.
Pulse Shift -7.5~3000μs
Alarm Real-time alarm signal indicated by horn or LED indicator
Real-time alarm signal indicated by horn or LED indicator
Accuracy of Calibrated Attenuator <±1dB/20dB
Probe Zero 0~999.9μs
Reject (0-99)%
RF Output Impedance 50Ω
Synchronous Data Collection and Output Mode Open Collector gate (OC gate))
Electric Noise Level <10%
Interface C6 transducer connections, USB HOST, LAN, RF OUT, Two-dimensional encoder interface
Power Supply Internal charger, AC Mains 220V; Large-capacity Li Battery; Battery Operating time: More than 6 hours
Environmental Temperature -30~50℃
Relative Humidity (20-95)%RH
Weight 1.4Kg (with battery)